Ghosts in the LIBS spectra acquired by an echelle type of spectrometer as a tool for the correction of wavelength calibration fluctuations during the analysis of large datasets
M. Horňáčková*, M. Veis, P. Veis
Department of Experimental Physics, Faculty of Mathematics Physics and Informatics, Comenius University, Bratislava, Slovakia
* Corresponding author: michaela.hornackova@fmph.uniba.sk

Echelle-type spectrometers are widely used in the field of LIBS (Laser-Induced Breakdown Spectroscopy) for atomic and molecular spectral acquisition due to their robust design, high spectral resolution, and broad wavelength coverage (typically 200-1000 nm) [1,2]. These features enable the simultaneous detection of numerous elemental emission lines. Owing to these advantages, LIBS systems equipped with an iCCD camera can be applied to a wide range of applications, including trace-element analysis in various matrices, while achieving very good limits of detection (LODs) [3].
However, in some cases, particularly when analysing samples containing very high concentrations of an element (e.g. substrate elements during the evaluation of LODs in liquids using surface-assisted (SA)-LIBS) or easily ionisable elements such as Na and K, undesirable effects may appear in LIBS spectra acquired with an echelle-type spectrometer. One of these effects is the occurrence of spectral "ghosts". Due to the design of an echelle spectrometer, which employs two dispersion elements (an echelle grating and a prism) arranged perpendicular to each other, the resulting echellogram forms a two-dimensional image of the spectrum in which each stripe corresponds to a particular diffraction order.
When the intensity of an emission line is extremely high, ghost images of this line may also appear in adjacent diffraction orders. As a result, additional spectral features arise in the measured spectrum that does not correspond to real emission lines. These ghost lines may overlap with spectral lines of interest and thus complicate the identification and quantification of elements present in the sample.
Proper calibration of the echelle spectrometer results in a symmetric distribution of ghost signals in adjacent diffraction orders, with approximately the same intensity. An asymmetric distribution of ghost signals indicates that the echellogram calibration is no longer valid. Temperature-induced shifts of the echellogram may lead to calibration errors and consequently to the displacement or even loss of selected spectral lines from their expected positions.
In this contribution, we propose the analysis of the intensity ratio of ghost signals in neighbouring diffraction orders as a tool for assessing the quality of spectrometer calibration. The approach is subsequently applied to the study of LODs, with and without the preselection of properly calibrated spectral lines.

[1] Shameem, K. M., Dhanada, V. S., Harikrishnan, S., George, S. D., Kartha, V. B., Santhosh, C., & Unnikrishnan, V. K., Echelle LIBS-Raman system: A versatile tool for mineralogical and archaeological applications (2020) Talanta, 208, 120482.
[2]Panne, U., Neuhauser, R. E., Theisen, M., Fink, H., & Niessner, R., Analysis of heavy metal aerosols on filters by laser-induced plasma spectroscopy (2001) Spectrochimica Acta Part B: Atomic Spectroscopy, 56, 839-850.
[3] Horňáčková, M., Plavčan, J., Horňáček, M., Hudec, P., Veis, P., Heavy Metals Detection in Zeolites Using the LIBS Method (2019) Atoms, 7(4), 98.
Acknowledgements The authors acknowledge the Scientific Grant Agency of the Slovak Republic (projects VEGA-1/0815/25, and VEGA-2/0120/25) and Slovak Research and Development Agency (projects APVV-22-0548, and APVV-23-0281) for financial support.

Pracovisko fakulty (katedra)/ Department of Faculty Department of experimental physics

Author

Michaela Hornackova (Department of experimental physics, Faculty of mathematics physics and informatics, Comenius University)

Co-authors

Dr Matej Veis (Department of experimental physics, Faculty of mathematics physics and informatics, Comenius University) Prof. Pavel Veis (Department of experimental physics, Faculty of mathematics physics and informatics, Comenius University)

Presentation materials

There are no materials yet.